Repository logo

Optimization of a RF Single Ion Paul Trap for a 88Sr+ Ion Optical Clock Comparison

Loading...
Thumbnail ImageThumbnail Image

Date

Journal Title

Journal ISSN

Volume Title

Publisher

Université d'Ottawa / University of Ottawa

Abstract

As part of the ongoing world-wide effort in improving time and frequency references, a high accuracy optical frequency standard was developed using the electric quadrupole allowed clock transition at 445 THz (674 nm) in a trapped and laser cooled 88Sr+ion. An ion trap system of the endcap design has been recently evaluated with a fractional frequency uncertainty which surpasses the accuracy of the current realization of the SI second. This thesis seeks to further evaluate the limiting systematic shifts of the device by optimizing a second ion trap reference based on a rf Paul trap design, which was then compared with the endcap trap reference frequency. The comparison of the two ion traps' reference frequencies confirmed an overall offset of -0.36 pm 0.08 Hz at the 445 THz reference frequency corresponding to a fractional frequency offset of 8 x 10^-16.

Description

Keywords

Optical Clock, Ion Trap, Frequency reference, Frequency standard, Time standard, Optical Clock comparison

Citation

Related Materials

Alternate Version